Functions and settings | 8
VERSION 1 | EN-US CIRUS
®
UPT-6400 Page 39 of 104
Menu
num-
ber
Menu
item
Available
only
when the
timer is
activated
(menu
number
401)
Description Setting range
312 Number
of
passes
through
current
trans-
former.
Number of conductors through the current transformer. In
this menu number, enter the number of times that the
secondary current line should pass through the current
transformer.
Refer to section Current transformer configuration [}29]
for further information.
1…9
313 Source
of AU-
TOCAL
tempera-
ture
Set the temperature to be used as the AUTOCAL tem-
perature.
• Internal (value
that the user en-
tered in menu
number 107)
• External (set-
point) (analog in-
put, terminals
36+39)
• External (CH1)
(terminals
39+40)
Machine menu
Menu
num-
ber
Menu
item
Available
only
when the
timer is
activated
(menu
number
401)
Description Setting range
401 Timer
function
Specify whether the temperature controller should work
with or without timer.
• ON
• ON, with start
monitoring
• OFF
402 Cooling
mode
x Configure the course of the cooling phase (end of cycle)
by selecting a type of cooling.
• Absolute: The cycle ends when the actual value falls
below the set cooling temperature.
• Relative: The cycle ends when the actual value falls
below the set proportional (percentage of setpoint)
cooling temperature.
• Time: The cycle ends when the cooling time has
elapsed.
• None: The cycle ends without defined cooling.
• Absolute (in °C)
• Relative (in %)
• Time (in sec)
• None
403 Start de-
lay
x Before starting the sealing time, this option can be used
to generate a start delay. The time ends when the start
signal is applied, and then the sealing time is activated.
0.00…9.98sec