3 – 28 HS40 Service Manual
■ Relative Focal Position
When modifying the 2D Depth value, determine whether or not to maintain a similar pre-
modification ratio of depth to focus position.
■ Enable MultiVision in DualLive Mode
In Dual Live Mode, irrespective of the Right Image settings, determine whether or not
MultiVision should apply to the Left Image. (In B/BC Dual Live Mode, regardless of the
settings, MultiVision will apply to both sides equally at all times.)
■ Enable ClearVision in DualLive Mode
In Dual Live Mode, irrespective of the Right Image settings, determine whether or not
ClearVision should apply to the Left Image. (In B/BC Dual Live Mode, regardless of the
settings, ClearVision will apply to both sides equally at all times.)
■ Preferred Probe Port
Select the Probe Port that takes the highest priority when the system starts or when you
select a probe.
■ Color Velocity Unit
Select the units of velocity in Color Mode.
■ Doppler Scale Unit
Select the units of measurement for the axis scale in Spectral Doppler Mode.
3.6.7.2. Display Setting
■ Show Image Info
Show or hide the image information. If the image information intrudes too much on the
screen, When it is turned off, it hides the image information.
■ Show Horizontal Scale Bar
Show or hide the Horizontal Scale Bar in the Scan UI.
■ Show Probe Center Line
Select whether or not to display the Center Line of the probe.
■ Display Tx Frequency
Specify how to display the Tx Frequency for each probe. If checked, the current state will be
indicated in MHz; if unchecked, the state will be indicated as PEN, GEN, or RES.
■ Display Linear Full Image in Single Mode
Show the full image of the Linear (including virtual convex) probe in Single Mode. When the
sides of an ultrasound image are not visible at a low depth, use this function to view the
entire image.