Chapter 9. Probes 9 23
9.2.2. Using the Biopsy Kit
WARNING
Only use the needles approved by the countrys government.
Verify the condition of the biopsy needle before use. Do not use a bent biopsy
needle.
The biopsy needle may bend during tissue penetration. The precise location
of the needle must be checked by monitoring the echo generated from the
needle.
Never use the biopsy kit to biopsy prostate tissue.
9.2.2.1. Before Using Biopsy Kit
WARNING
Do not attempt to use the biopsy kit until you have read the instructions for
installing the sheath and verifying alignment of the needle guide.
Always ensure that the probe and the needle guide are secured on both the
left and the right.
Do not use in IVF, CVS, or PUBS procedures.
Ultrasonographic scanning using the biopsy kit must be conducted by medical doctors or
experienced medical staff with appropriate qualifications. Always, without fail, verify all safety
procedures and disinfection.
Biopsy kits not supplied by Samsung Medison may fail to fit the probe properly. Always use
biopsy kits supplied by Samsung Medison, since improper setup may adversely affect the
patient.
Inspect all components. Ensure that the biopsy kit you are using is the correct one for the probe,
the system, and system software.