EasyManua.ls Logo

Samsung HS40 - Detailed Inspections

Samsung HS40
308 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4. Product Inspection 4 – 3
Patient
Enter information in Patient and check that the same Patient information is displayed in the
Report, SonoView, etc.
End Ex am
Make an arbitrary measurement in New Patient and see whether the measurement is erased
when you press End Exam.
Probe Key
Perform a Probe Change and check that it functions correctly.
4.1.2. Detailed Inspections
2D Mode
1. Perform a Knife Test to check that a part of image is not displayed.
2. Perform a Phantom test to check for errors in the image.
3. Check the Freeze Cine actions (broken image, Auto Run, Auto Run Speed, Trackball
Cine).
4. Check for changes in the image brightness when you adjust the Gain.
5. Check that the TGC Gain operates correctly by adjusting it and checking for changes in
the image brightness in accordance with depth.
6. Perform Left/Right Flip, Up/Down Direction, and Rotation to check whether the image
changes its orientation correctly.
7. Check whether or not the Select Image menus (EE, DR, View Area, Tissue, Frame Rate)
work.
8. Check for errors in frequency (Phantom, Res, Pen, Gen).
9. Check for changes in the image in accordance with changes in depth.
10. Check for changes in the image by depth when you change the focus.
11. Check whether or not the Image Compensation Mode works.

Table of Contents