9 – 30 HS40 Service Manual
9.2.4. Cleaning and Disinfecting the Biopsy Kit
Always use protective eyewear and gloves when cleaning and disinfecting the biopsy
kit.
Wash and disinfect the biopsy kit to reduce pathogens to the level of 10
-6
.
Reusable Biopsy adapters must be cleaned and disinfected after each use. Disposable Biopsy
adapters and disposable Biopsy kit parts need to be disposed as medical waste after each use.
Cleaning is an important procedure that must be carried out before disinfecting or sterilizing
reusable Biopsy adapters. Read the biopsy kit user manual carefully before cleaning and
disinfecting.
9.2.4.1. Cleaning Biopsy Kit
1. After use, the biopsy kit is removed from the probe.
2. Disassemble the biopsy kit into its component parts and discard the disposable parts in the
clinical waste bin.
3. Rinse with water to remove excess soil from device.
4. Prepare an enzymatic solution (e.g. Enzol), following manufacturer instructions for proper
dilution and immersion time. Immerse the parts in the detergent and allow soaking for the
prescribed time.
5. Brush the parts thoroughly using a soft bristle brush. Clean the lumen using an appropriate
sized brush. If any visible parts remain, continue brushing until clean.
6. Thoroughly rinse the device with water and dry the parts. Discard the used soft cloth to
clinical waste.