Chapter 9. Probes 9 31
9.2.4.2. Disinfecting the Biopsy Kit
CAUTION
Always refer to the biopsy kit user manual for the latest information on the
supported methods and disinfectants.
A stainless steel biopsy adapter is reusable; a plastic biopsy adapter may be
reusable. A plastic biopsy needle guide is in general disposable.
Plastic biopsy kits can only be disinfected by using a chemically compatible
cold-disinfectant, either manual or with a medical washer-disinfector. Do not
use an autoclave or sterilize using gas.
Please refer to the user manual of the disinfectant for storage, use, and disposal of the
disinfectant.
1. Re-assemble the reusable parts of the biopsy adapter, if applicable.
2. Disinfect the biopsy adapter for the duration and temperature as specified in the user
instructions of the adapter.
3. After disinfection, follow the proper post-disinfection procedure for the disinfection method
used. (Refer to the instructions for the disinfectant, disinfecting equipment, etc.)
4. Inspect the components for damage such as cracks, rust or breakage and if damage is
discovered stop using the device and contact the Samsung Medison Customer Support
Department.
5. Properly pack the disinfected adapter and label it according to local regulations.
9.2.4.3. Sterilizing the Stainless Steel Biopsy Kit
1. Re-assemble the reusable parts of the stainless steel biopsy adapter, if applicable.
2. Sterilize the stainless steel adapter by autoclaving (Steam) or using gas (Ethylene Oxide).
3. After sterilization, follow the proper post-sterilization procedure for the sterilization method
used. (Refer to the instruction of the sterilization equipment.)
4. Inspect the components for damage such as cracks, rust or breakage and if damage is
discovered stop using the device and contact the Samsung Medison Customer Support
Department.
5. Properly pack the sterilized adapter and label it according to local regulations.