Chapter 3. Installing Product 3 – 13
3.5. System Power
Boot up the system for use.
Make sure to connect the probes and peripheral devices that will be used before
powering on the system. If you attempt to connect them during system use, it
may lead to patient injury or irreparable damage to the console.
3.5.1. Turning the Power On
Press the On/Off button when the power is off. Booting begins, and the product logo appears on
the screen. When booting is completed, the 2D mode screen will appear in End Exam status.
Before starting the diagnosis, you must register the pati
ent information.
NOTE
If the power switch near the power connection port on the rear panel of the
product has been switched off, wait for 10 seconds before turning on the
product.
Do not press keyboard keys or buttons while booting is in progress. Doing so
may cause the system to malfunction.
If you turn on the power after turning it off suddenly, the system may turn on
and off momentarily. This is a feature of the built-
in Intel PC Main Board of the
product, and does not indicate a system error.
3.5.2. Shutting Down the System
Press the On/Off button while using the system to initiate shutdown. Click Shut Down on the
screen to shut down the system, or click Cancel to cancel.
CAUTION
If the system does not turn off, keep the On/Off button depressed for five
seconds or longer to force a shutdown.
If the ultrasound system is unplugged while working or turned off by forcibly
shutting down, the files may get corrupted and it may result in system failure
or data loss.
To ensure that the product is safely cut off from electrical power, set the
power switch at the rear of the product to the Off position after using the
product.
Press the On/Off button twice to turn off the system. The Select Next
Operation window appears when you press the On/Off button once. The
system will turn off if you press the button again.