9 – 26 HS40 Service Manual
9.2.3. Assembling the Biopsy Kit
9.2.3.1. Stainless Steel Biopsy Kit
1. Place a Sheath up to the top of the probe’s handle.
2. Mount the biopsy adapter onto the probe. If the surface of probe is fluted, mount the adapter
in accordance with it.
3. Install the needle into the needle guide and start the exam.