Chapter 2. Safety 2 9
WARNING
Electric shock may result if this system, in
cluding all of its externally mounted
recording and monitoring devices, is not properly grounded.
Never open the cover of the product. Hazardous voltages are present inside.
All internal adjustments and replacements must be made by qualified
Samsung Medison Customer Support Department personnel.
Always check the products housing, cables, cords, and plugs before using
the product. Disconnect the power source and do not use the equipment if
the housing is damaged such as cracked, and chipped, or if the cable is
worn.
Always disconnect the system from the wall outlet prior to cleaning the
system.
All patient contact devices, such as probes and ECG leads, must be
removed from the patient prior to application of a high voltage defibrillation
pulse.
The use of flammable anesthetic gas or oxidizing gases (N
2
O) should be
avoided. There is a risk of explosion.
Avoid installing the system in such a way that it is difficult for the operator to
disconnect it from the power source.
Do not use together with HF surgical eq
uipment. HF surgical equipment may
be damaged, which may result in fire.
The system must only be connected to a supply mains with protective earth
to avoid risk of electric shock.
CAUTION
The system has been designed for 100-240VAC; you should select the input
voltage of any connected printer and VCR. Prior to connecting a peripheral
power cord, verify that the voltage indicated on the power cord matches the
voltage rating of the peripheral device.
An isolation transformer protects the system from power surges. The
isolation transformer continues to operate when the system is in standby.
Do not immerse the cable in liquids. Cables are not waterproof.
Make sure that the inside of the system is not exposed to or flooded with
liquids. In such cases, fire, electric shock, injury, or damage to the product
may occur.
The auxiliary socket outlets installed on this system are rated 100-240VAC
with maximum total load of 150VA. Use these outlets only for supplying
power to equipment that is intended to be part of the ultrasound system. Do
not connect additional multiple-socket outlets or extension cords to the
system.
Do not connect any peripheral devices that are not listed in this manual to
the auxiliary socket outlet of the system. It may cause an electrical hazard.
Do not touch SIP/SOP and the patient simultaneously. There is a risk of
electric shock from leakage current.