2 – 12 HS40 Service Manual
2.3.5.2. Approved Cables, Probes and Peripherals for EMC
■ Cables
Cables connected to this product may affect its emissions; use only the cable types and
lengths listed in the table below.
Cable Type Length
VGA Shielded Normal
USB Shielded Normal
LAN(RJ45) Twisted pair Any
S-Video Shielded Normal
Foot Switch Shielded 2.99m
Audio R.L Shielded Normal
Parallel Shielded Normal
HDMI Shielded Normal
ECG 3-lead Shielded Normal
■ Probes
The image probe used with this product may affect its emission. The probe listed in ‘Chapter
9. Probes’ when used with this product, have been tested to comply with the group1 Class A
emission as required by International Standard CISPR 11.
■ Peripherals
Peripherals used with this product may affect its emissions.
When connecting other customer
-supplied accessories to the system, it is the
user’s responsibility to ensure the electromagnetic compatibility of the system.
The use of cables, probes, and peripherals other than those specified may result
in increased emission or decreased Immunity of the Ultra
sound System.