3 16 HS40 Service Manual
3.6.2.2. Date and Time
Set the date and time for use in the system.
CAUTION
Before starting diagnosis, please confirm the time, date and time zone and
modify the settings if necessary.
Make sure to set the date and time to the time zone in which the system is
installed. If it is not set, the date and time can be automatically synchronized
to the time zone of the country of the manufacturer.
NOTE
You cannot change the date and time when a patient ID has been
registered. To change the date and time, you should finish the current
diagnosis by pressing the End Ex am button on the control panel.
When the system time and date are changed, the change is not reflected in
the time and date indicated in previous diagnosis results.
Date Format
Specify the date format. Select a date format using the combo button. The date format that
you specify will be applied to various date fields in Patient Information.
Time Format
Specify the time format. Press the combo button to select the preferred display format (12
Hour or 24 Hour).
Date and Time
Select settings for date and time. When you press this button, the Date and Time screen
appears.
1. Set the date and time by using the trackball and the Set button on the control panel.
2. Press Save to apply the settings. Tap Cancel or press Ex it button on the control panel to
cancel.
Time Zone
Specify the local time zone where the system is used. When you press this button, the Time
Zone Settings screen appears.
1. Press Time Zone.
2. Specify the time zone by using the trackball and the Set button for the combo button.
3. Press OK to apply the settings. Tap Cancel to cancel.