t Define how the ports should be activated depending on the check results
*Ports not active
Within checkweighing range
: Ports are only activated when the results
are within the checkweighing range.
Ports always activated: The ports are always activated with each weight value.
When stable and in checkweighing range: The ports are only activated
as soon as the balance becomes stable and the results are in the checkweighing
range.
At stability: The ports are always activated as soon as the balance has stabilized.
Once at stability and in the checkweighing range: The ports are activated
on a one-time basis (not dynamically) as soon as the balance has stabilized
and the results are in the checkweighing range; effective until the balance has been
unloaded.
t Define whether or not the automatic counter for positive results should
be activated.
*No automatic counter
Counter for OK values:
All positive checked results are counted.
y An overview of all settings is displayed.
t Check all settings and change as required.
y A message asks whether you would like to add another application.
t If required, select additional applications to be combined.
y Now you will be prompted to configure the weighing and printing functions.
t Check all settings and change as required.
t Enter a short name and a description for the new task.
t To save the new task, select Save.
Checkweighing
110 Cubis MSU User Manual