EasyManuals Logo

Sartorius Cubis Series User Manual

Sartorius Cubis Series
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
Interfaces
Cubis MSU User Manual 147
SICS Commands
The computer connected via the data port can send control commands
to the balance to control balance and application program functions.
h
Please consult the separate SICS documentation for a description of the syntax.
1)
Some labeled commands can also be mini-SICS commands.
Level 0 Version 2.3x
Command Action/Function
@ Reset all SICS commands
I0 List all available commands
I1 Send information about the level and its versions
I2 Balance model prompt
I3 Prompt for input of software version of balance (BAC)
I4 Prompt for balance serial number
I5 Prompt for software version of display (APC)
S
1)
Send weight value at stability
SI
1)
Send weight value without stability
SIR
1)
Send automatic weight values at and without stability
Z Zero the balance at stability
ZI Zero the balance without stability
Level 1 Version 2.2x
Command Action/Function
T
1)
Tare the balance at stability
TI
1)
Tare the balance without stability
D Write text in display
DW Delete text from display
K Key control
TA Prompt for and allocate tare memory
TAC Delete tare memory
SR Send weight value if weight changes
Level 2
Command Action/Function
M13 Activate/deactivate touch screen softkeys
PWR Turn balance on/off (standby)
Remote control (remote)
Command Action/Function
P112 Write text in selected line on display
P113 Delete text from selected line in display
P114 Overwrite task or user names
P120 Turn off bar graph in checkweigher
P121 Turn on bar graph in checkweigher
RM20 Activate/deactivate user input
RM30 Assign new function to softkeys
RM36 Assign/prompt for function of multiple softkey lines
RM38 Activate RM36-assigned softkey lines
RM39 Activate/deactivate RM30-assigned softkey functions
Additional Sartorius commands
Command Action/Function
SA
1)
Send weight value at stability and store in alibi memory (with optional label)
CMD
1)
Execute application command
PAR
1)
Query parameter

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model