Cubis MSU User Manual 165
Specifications
Model-specific Data
Precision balances
Models MSU623S MSU623P MSU323S
Readability mg 1 1/2/5 1
Weighing capacity g 620 150/300/620 320
Tare range (subtractive) g – 620 – 620 – 320
Repeatability <±mg 0.7 1/2/4 0.7
Linearity <±mg 2 5 2
Corner load (test load [g]) mg 2 (200) 4 (200) 2 (200)
Min. initial weight* g 1.5 1.5 1.5
Sensitivity drift between +10 to +30°C ±ppm/K2 2 2
Typical stabilization time s < 0.8 < 0.8 < 0.8
Typical measurement time s < 1 < 1 < 1
External standard calibration value g 500 (E2) 500 (F1) 200 (E2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W + D) mm 140 +140
Weighing chamber height mm 172
(draft shield DU)
Protection Protected against dust and water
* = Position according to OIML R76
Models MSU14202S MSU14202P MSU10202S MSU8202S
Readability mg 10 10/20/50 10 10
Weighing capacity g 14,200 3500/7000/14,200 10,200 8200
Tare range (subtractive) g - 14,200 - 14,200 – 10,200 – 8200
Repeatability <±mg 10 10/20/40 7 7
Linearity <±mg 30 50 20 20
Corner load (test load [g]) mg 20 (5000) 40 (5000) 20 (5000) 20 (5000)
Min. initial weight* g 15 15 12 12
Sensitivity drift between +10 to +30°C ±ppm/K1.5 1.5 2 2
Typical stabilization time s < 1 < 1 < 1 < 1
Typical measurement time s < 1.5 < 1.5 < 1.5 < 1.5
External standard calibration value kg 10 (E2) 10 (E2) 10 (E2) 5 (E2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W + D) mm 206 +206
Protection Protected against dust and water
* = Position according to OIML R76