170 Cubis MSU User Manual
Specifications
Model-specific Data
Verified models with EC Type-Approval Certificate: Precision balances
Models MSU6202S- MSU6202P- MSU5202S- MSU4202S-
0CE 0CE 0CE 0CE
Accuracy class* KK k K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* g 0.01 0.01/0.02/0.05 0.01 0.01
Weighing capacity max* g 6200 1500/3000/ 5200 4200
6200
Calibration value e* g 0.1 0.1 0.1 0.1
Min. load min* g 0.5 0.5 1 0.5
Tare equalization range (subtractive) < 100% from max. weighing capacity
Application range according to DIR* g 0.5 – 6200 0.5 – 6200 1 – 5200 0.5 – 4200
Min. initial weight** g 12 12 10 12
Typical stabilization time s < 1 < 1 < 0.8 < 0.8
Typical measurement time s < 1.5 < 1.5 < 1 < 1
Application range (temperature):
With “isoCAL” function +5 to +40°C +5 to +40°C +5 to +40°C
Without “isoCAL” function +10 to +30°C +10 to +30°C +10 to +30°C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W + D) mm 206 +206
Protection Protected against dust and water
Models MSU2202S- MSU1202S- MSU12201S- MSU8201S- MSU5201S-
0CE 0CE 0CE 0CE 0CE
Accuracy class* KKKKK
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 10 10 100 100 100
Weighing capacity max* g 2200 1200 12,200 8200 5200
Calibration value e* g 0.1 0.1 1 1 1
Min. load min* g 0.5 0.5 5 5 5
Tare equalization range (subtractive) < 100% from max. weighing capacity
Application range according to DIR* g 0.5 – 2200 0.5 – 1200 5 – 12,200 5 – 8200 5 – 5200
Min. initial weight** g 12 12 100 100 100
Typical stabilization time s < 0.8 < 0.8 < 0.8 < 0.8 < 0.8
Typical measurement time s < 1 < 1 < 1 < 1 < 1
External standard calibration value kg 2 (F1) 1 (F1) 10 (F1) 5 (F2) 5 (E2)
(min. accuracy class)
Application range (temperature)
With “isoCAL” function: +5 to +40°C | Without “isoCAL” function: +10 to +30°C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W + D) mm 206 +206
Protection Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model