Leveling the Balance, Setting the Level Indicator
h
Leveling the balance compensates for slant or unevenness at the place
of installation. The balance must be perfectly horizontal to ensure con-
sistent, reproducible weighing results. All models are equipped with
an electronic tilt angle detection feature. If the balance is not level,
all balance models issue an alert on the display.
Push-button automatic leveling is available on models with motorized
leveling feet. For models with manual leveling feet, please follow
the instructions on the display.
Leveling the Balance Manually
t The balance is leveled using both front leveling feet.
t Screw in both back leveling feet (only for models with back leveling feet).
t Turn the two front feet as shown in the illustration until the air bubble is cen-
tered within the circle of the level indicator.
y Normally, several leveling steps are required.
t Turn both back leveling feet until they touch the setup surface
(only for models with back leveling feet).
Warm-up Time
y To return precise results, the balance must warm up for at least 30 minutes
after initial connection to the power supply.
Only after this time will the device have reached the required operating temper-
ature.
When a verified balance used in legal metrology (legal-for-trade applica-
tions) is connected to the power, it must warm up for at least two hours
before operation.
t Turn on the devices via the A key. The following appears on the display:
y The operator can now follow the brief instructions to configure the balance
settings before beginning weighing operations.