System Settings (Menu)
Cubis MSU User Manual 45
Leveler Configuration (Q-Level)
You can configure the leveling function here.
Leveling Notice
Off, no indication: Leveling prompt is disabled.
Display status: As soon as leveling is required,
Level me! appears and flashes.
Alarm message, leveling required: As soon as leveling is required,
Level me! appears and flashes. If this notice is ignored, an alarm message appears
after five minutes. If this occurs, you can only continue working after the balance
has been leveled.
Automatic Leveling (Balances with Motorized Leveling Feet Only)
Switch off, start with key: Select Start to activate automatic leveling.
Switch on fully automatic: The balance is leveled automatically before every
calibration/adjustment routine.
t To save the settings, select Save.
Power On Mode
t Select the balance’s startup mode for when the balance is switched on or when
the on/off key A is pressed repeatedly.
Options: On/standby or Always on
t To save the settings, select Save.
Power Saving Mode
t Switch the power saving mode of the balance on or off.
Off, no function: Power saving mode is not in use.
Automatic shutoff after 2 minutes: The backlight on the display
and control unit switches off if the device is idle for twominutes.
t To save the settings, select Save.
Restore Factory Settings
You can reset all balance settings back to the default delivery status here
(after confirming the reset when prompted).