System Settings (Menu)
46 Cubis MSU User Manual
Configure Data Output (Print Parameters)
Here you can configure all settings for printing and data output.
Several of these settings depend on the interface configuration
(see Section “Interfaces”).
h
Factory settings are marked with an asterisk (*).
In order to view detailed information about individual settings, select Info.
Printout interface: Options: COM A, B, C (file/SD memory card), D
Protocol: Options: Print / SBI / xBPI / Web services / SICS / Second display.
This setting cannot be changed here; it is defined when configuring the interfaces.
h
The parameters and settings that are displayed here depend on the configured
protocol.
GLP printing: When should ISO/GLP printing be allowed? Options:
*Off / For calibration/adjustment only / Always on.
Taring after printing: Automatic tare after print
command. Options: *Off / On
Print event: What event should trigger printing?
Options: *PRINT key / Starting a task / Initialization of an application / Application
result / Printing of component / Result from an application (evaluation of applica-
tion) / End of task / Calibration/adjustment result
Print key elements: Which elements should be printed?
h
For further information, please refer to Configuring Printouts in Section
Interfaces.
Set Date/Time
t Enter the current date and time in the prescribed format.
Configure Display and Control Unit Properties
The following properties of the display and control unit can be changed as neces-
sary: Language, Background color, Brightness, Acoustic signal.
These general properties are applied to all user profiles, are also used when no user
is logged in, and serve as a template when a new user profile is created.