t Enter the range of instability for the start of the measurement:
0.1%, 0.2%, 0.5% ... *10% ... 50% from measured object
This entry estimates the strength of the instability that is caused by the activity
(e. g., animal movement). You can use the rough estimation or a more detailed
estimation using percentages.
The weight value changes with each movement of the weighing pan. As soon
as three consecutive measurement values are within the selected instability range,
the interval measurement begins.
Example: You are weighing animals with an average weight of approx. 300 g.
With the “5.0% of object” setting, the measurement starts when the movements
are within a range of 15 g, i.e., ± 7.5 g. With the “50% of object” setting, the
interval measurement would start when the weight value was within a range
of 150 g, i.e. ± 75 g.
t Define the number of measurements to be used for the average value.
Enter fixed number: You must now enter a fixed number.
*Number can be changed during process: You will define the number when
you execute the task.
Aktivität
Anzeigeschritte
Messung bei 5% ± 7,5 g
Start bei Mindestlast 100 Anzeigeschritte
Zeit
Messung bei 50% ±75 g
400
350
300
250
200
150
100
50
0
86 Cubis MSU User Manual
Averaging

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model