Cubis MSU User Manual 95
Weighing in Percent
t Define which percent calculation should be used.
Example: Weight value = 10 g, reference weight = 50 g
*Residue: The result is the percentage share of the weight value of the reference
weight (residue: weight value / reference weight * 100%; in this example: 20%)
Loss: The result is the percentage loss between the weight value and the reference
weight (loss: weight value - reference weight / reference weight * 100%; in this
example: -80%)
Ratio 1 (DR): The result is the percentage ratio of the difference to the weight
value (ratio1: reference weight - weight value / weight value * 100%; in this
example: 400%)
Ratio 2 (OR): The result is the percentage ratio of the reference weight to the
weight value (ratio2: reference weight / weight value * 100%; in the example:
500%)
t Define the number of decimal places to be used for the display of the
percentage.
none ... 1, *2 , 3, 4
t Define the accuracy to be used to store the weight value for the reference
weight.
*Normal resolution (display)
10-fold (+ 1 dec. place)
100-fold (+2 dec. places)
y An overview of all settings is displayed.
t Check all settings and change as required.
y A message asks whether you would like to add another application.
t If required, select additional applications to be combined.
y Now you will be prompted to configure the weighing and printing functions.
t Check all settings and change as required.
t Enter a short name and a description for the new task.
t To save the new task, select Save.

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model

Related product manuals