HF Series Generators
T roubleshooting
TR-1005R4
33
ERROR CODE : E06
DESCRIPTION : Prep or/and Exposure orders activated during power-up routine.
ERROR TYPE : Informative. Generator re-start automatically once error is solved.
APPLICABLE TO : All Generators
APPEARS WHEN : At any moment once initialization phase is over.
INFORMATION / SYMPTOM : This Error Code needs to turn OFF/ON the Generator to be fixed.
It has not been possible to establish a correct communication between ATP Console and HT
Controller Boards.
POSIBLE CAUSES
The Generator has been detected Prep”or“Exposure signals during initialization.
ACTIONS
1. In Generators equipped with “Prep and Exposure” keys on the Console, check that not object
is activating the function. Also check for possible damages on keys.
2. Check damages on flat cable with connector J9.
3. In Generators with external Handswitch, check contacts, cable and connectors.
4. If the error remains, check continuity between TS1-37 and J2-17 in ATP Console Board and
TS1-36 and J2-4 also in ATP Console Board.
5. If not solved yet, replace the ATP Console Board.
ERROR CODE : E07
DESCRIPTION : Wrong data for X-ray Tube-2.
ERROR TYPE : Fatal error. Generator can not continue with power up.
APPLICABLE TO : All Generators
APPEARS WHEN : Only during initialization phase.
INFORMATION / SYMPTOM : This Error Code needs to turn OFF/ON the Generator to be fixed.
It has not been possible to establish a correct communication between ATP Console and HT
Controller Boards during power ON.
POSIBLE CAUSES
Wrong configuration of X-ray Tube-2.
Corrupted calibration data.
Data on Extended Memory Location E18 are larger than the maximum allowed.
ACTIONS
1. Check data in Extended Memory Location E18.
2. If it is OK, replace U3 in HT Controller Board.