HF Series Generators
T roubleshooting
TR-1005R4
32
ERROR CODE : E04
DESCRIPTION : “Prep” signal received without Console order.
ERROR TYPE : Informative. Generator re-start automatically once error is solved.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment once initialization phase is over.
INFORMATION / SYMPTOM : This Error Code needs to turn OFF/ON the Generator to be fixed.
It has not been possible to establish a correct communication between ATP Console and HT
Controller Boards.
POSIBLE CAUSES
The Generator receives the “Prep” signal without a Console command.
ACTIONS
1. Check continuity between P1-3 in HT Controller Board and J3-4 in ATP Console Board. Check
with special care connector 6J3 of the Generator.
2. If all is correct, check signal from P1-3 to U5 in HT Controller Board.
3. If the signal is OK, replace HT Controller Board.
ERROR CODE : E05
DESCRIPTION : “Fluoro” signal active without request.
ERROR TYPE : Fatal error. Generator can not continue with Power up.
APPLICABLE TO : R&F Generators or Generators configured by mistake as R&F.
APPEARS WHEN : Only during initialization phase.
INFORMATION / SYMPTOM : This Error Code needs to turn OFF/ON the Generator to be fixed.
It has not been possible to establish a correct communication between ATP Console and Fluoro CPU
Boards during power ON and the Generator is not able to make exposures.
POSIBLE CAUSES
The Generator receives the “Fluoro” signal without Console command.
“Exp” has been pressed in a Workstation not configured for standard R&F.
ACTIONS
1. Check an unwilling press on the Fluoro pedal.
2. If not the case, check continuity between TS1-37 and J2-17 in ATP Console Board.
3. Then, check signal -FL EXP in ATP Console Board.
4. If the Signal is OK, replace the ATP Console Board.