HF Series Generators
T roubleshooting
TR-1005R4
31
ERROR CODE : E02
DESCRIPTION : Failure in power up routine. RAD Generator configured as R&F Generator.
No communication between ATP Console CPU Board and Fluoro CPU Board.
ERROR TYPE : Fatal error. Generator can not continue with power up.
APPLICABLE TO : R&F Generators or Generators configured by mistake as R&F.
APPEARS WHEN : Only during initialization phase. If it appears during normal functioning of equipment, it means that
a problem has caused a power off in the Console.
INFORMATION / SYMPTOM : This Error Code needs to turn OFF/ON the Generator to be fixed.
It has not been possible to establish a correct communication between ATP Console and Fluoro CPU
Boards during power ON and the Generator is not able to make exposures.
POSIBLE CAUSES
A RAD Generator has been configured by mistake as a R&F Generator. See Configuration Section.
The communication between ATP Console Board and Fluoro CPU Board is not reliable.
The Fluoro CPU Board does not work properly.
ACTIONS
1. Check Communication Cable between J4 of ATP Console and J4 of Fluoro CPU Boards.
2. Check and replace if necessary U12 in Fluoro CPU Board.
3. If Fluoro values are not displayed on the Console, check DC supplies and Fluoro CPU Board.
Replace Fluoro CPU Board if necessary .
ERROR CODE : E03
DESCRIPTION : Failure in power up routine. All workstation are unconfigured.
ERROR TYPE : Fatal error. Generator can not continue with power up.
APPLICABLE TO : All Generators.
APPEARS WHEN : Only during initialization phase. If it appears during normal functioning of equipment, it means that
a problem has caused a power off in the Console.
INFORMATION / SYMPTOM : This Error Code needs to turn OFF/ON the Generator to be fixed.
It has not been possible to establish a correct communication between ATP Console and HT
Controller Boards during power ON and the Generator is not able to make Exposures.
POSIBLE CAUSES
All workstations have been configured as tube=0.
The EEPROM (U18) in ATP Console Board is defective.
The ATP Console Board is not able to communicate with U18.
ACTIONS
1. Configure the workstations as referred in Service Manual.
2. Change U18 and configure the workstations as referred in Service Manual.
3. If not fixed after previous steps, replace ATP Console Board and configure the workstations.