HF Series Generators
T roubleshooting
TR-1005R4
55
ERROR CODE : E23
DESCRIPTION : Calibration data not stored
ERROR TYPE : Indicative although it is almost impossible to find if not provoked.
APPLICABLE TO : All Generators.
APPEARS WHEN : When trying to save a calibration data.
INFORMATION / SYMPTOM : Calibration data not stored. Calibration value intended to be stored has not been properly
recorded.
POSIBLE CAUSES
This problem may be shown randomly.
This problem is continuos or occurs frequently and the communication between the Generator
Cabinet and the Console is too noisy .
ACTIONS
1. Store data again.
2. Check communication cables.
3. Route communication cables in different way.