HF Series Generators
T roubleshooting
TR-1005R4
56
ERROR CODE : E24
DESCRIPTION : The Bucky has not been detected to be moving.
ERROR TYPE : Indicative. Does not allow exposition.
APPLICABLE TO : All Generators.
APPEARS WHEN : At the beginning of exposition.
INFORMATION / SYMPTOM : Wrong Exposure signal from Buckies.
POSIBLE CAUSES
Malfunction of the bucky.
Wrong wiring of the bucky.
Wrong bucky selection.
ACTIONS
CHECK
-- DRCMA AT PREP
IS IT AT LOW
LEVEL?
NO
CHECK K2 FOR
BUCKY1&K3FOR
BUCKY 2
ARETHEYON?
YES
REPAIR
CONNECTION
CHECK-BUCKY
MOTION 1&2 AT
INTERFACE BOARD
P4-11 & P4-10 AT LOW
LEVEL IS IT OK?
YES
NO
PROBLEM AT BUCKY
REPLACE
INTERFACE BOARD
REPLACE
ATP CONSOLE BOARD
YES
CHECK AND REPAIR
LINK J3--14 //P4-12 AT
INTERFACE BOARD
J3-- 11 / P4-13 AT
INTERFACE BOARD
NO
NO
YES
REVIEW WORK STATION
CONFIGURATION
IS IT OK?
CONFIGURE
CORRECTLY
YES
NO
CHECK
J3-14,J3-11 AT
ATP CONS.
IS IT OK?