HF Series Generators
T roubleshooting
TR-1005R4
60
ERROR CODE : E29
DESCRIPTION : “ALL CLEAR” signal not active.
ERROR TYPE : Indicative. Does not allow “Prep”.
APPLICABLE TO : Nucletron Generators.
APPEARS WHEN : When pressing Prep” in P00 mode.
INFORMATION / SYMPTOM : Exposure is not allowed.
POSIBLE CAUSES
“All Clear” signal is not ready after 5 seconds with “Prep” activated.
ACTIONS
Release the exposure controls, press the System Reset” button and check the system.
ERROR CODE : E30
DESCRIPTION : Relay K1 not active.
ERROR TYPE : Indicative. Does not allow “Prep”.
APPLICABLE TO : Nucletron Generators.
APPEARS WHEN : When pressing Prep” in P00 or P03 mode.
INFORMATION / SYMPTOM : Test--scan not started.
POSIBLE CAUSES
MOSTA signal is not received at system or relay K1 is not active.
ACTIONS
Release the exposure controls, press the System Reset” push-button and check the system.