HF Series Generators
T roubleshooting
TR-1005R4
68
ERROR CODE : E48
DESCRIPTION : Collimator Error .
ERROR TYPE : Informative. Does not allow exposure.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : Showed as “Technique Error” in Consoles with this light indicator. For the rest of Consoles E48
appears displayed.
POSIBLE CAUSES
Collimator blades closed or in motion during exposure.
Defective Collimator.
ACTIONS
1. Check collimator blades.
2. If it is OK, check the TS3-20 is at low level.
• If it is not at low level : Problem in Collimator .
• If signal at low level, check J2-6 in ATP Console
G if it is at low level, replace ATP Console Board
G If is at high level, check and fix link between J2-6 and TS3-20.
ERROR CODE : E49
DESCRIPTION : Exposure Cycle Error.
ERROR TYPE : Informative. Does not allow exposure.
APPLICABLE TO : Generators with two X-ray Tubes for RAD only with “Exposure Cycle” feature (a.e. Brandis).
APPEARS WHEN : Exposure is not performed or it is aborted.
INFORMATION / SYMPTOM : E49 appears on Console.
POSIBLE CAUSES
Operator releases “EXP” button before exposure time has ended.
Exposure order is interrupted.
ACTIONS
1. Press the respective button on the Console to reset the Error indication.
2. Repeat the exposure.
3. If error remains, check Handswitch or “EXP” button and replace what is wrong.