HF Series Generators
T roubleshooting
TR-1005R4
69
ERROR CODE : E50
DESCRIPTION : Interrupted Exposure.
ERROR TYPE : Indicative.
APPLICABLE TO : All Generators.
APPEARS WHEN : When operator releases “EXP” button befo re exposure time has ended.
INFORMATION / SYMPTOM : E50 appears on Console.
POSIBLE CAUSES
Operator releases “EXP” button before exposure time has ended.
ACTIONS
1. Press the respective button on the Console to reset the Error indication.
2. Repeat the exposure.
3. If error remains, check Handswitch or “EXP” button and replace what is wrong.
ERROR CODE : E51
DESCRIPTION : Checksum failure or EPROM corrupted.
ERROR TYPE : Indicative. System does not allow exposition.
APPLICABLE TO : Generators with DRAC or LV-DRAC.
APPEARS WHEN : After self-test.
INFORMATION / SYMPTOM : At power On, after the Generator autocheck, E51 is displayed and it is not possible to make
Exposures.
POSIBLE CAUSES
DRAC program memory corrupted.
ACTIONS
1. Replace U17 in Control DRAC Board.