EasyManua.ls Logo

Sedecal SHF-310 - Page 309

Default Icon
427 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HF Series Generators
T roubleshooting
TR-1005R4
69
ERROR CODE : E50
DESCRIPTION : Interrupted Exposure.
ERROR TYPE : Indicative.
APPLICABLE TO : All Generators.
APPEARS WHEN : When operator releases EXP” button befo re exposure time has ended.
INFORMATION / SYMPTOM : E50 appears on Console.
POSIBLE CAUSES
Operator releases “EXP” button before exposure time has ended.
ACTIONS
1. Press the respective button on the Console to reset the Error indication.
2. Repeat the exposure.
3. If error remains, check Handswitch or EXP” button and replace what is wrong.
ERROR CODE : E51
DESCRIPTION : Checksum failure or EPROM corrupted.
ERROR TYPE : Indicative. System does not allow exposition.
APPLICABLE TO : Generators with DRAC or LV-DRAC.
APPEARS WHEN : After self-test.
INFORMATION / SYMPTOM : At power On, after the Generator autocheck, E51 is displayed and it is not possible to make
Exposures.
POSIBLE CAUSES
DRAC program memory corrupted.
ACTIONS
1. Replace U17 in Control DRAC Board.