HF Series Generators
Troubleshooting
TR-1005R4
109
5.2 FILAMENT CONTROL BOARD
JUMPERS / SWITCHES
JUMPER / SWITCH DESCRIPTION
JP1 A A3004-- 05 VERSION
B A3004--04 VERSION
BOARD CONFIGURATION. JP1-A FOR NORMAL APPLICATION
LEDs
LED COLOR DESCRIPTION
DS1 GREEN
FILAMENT DRIVER STATUS INDICATOR:
DS2 GREEN
BOTH LEDs
RE OFF: 220
C POWER SUPPL
IS MISSING
ONLY ONE LED IS OFF: THE BOARD IS DEFECTIVE
5.3 INTERFACE CONTROL BOARD
JUMPERS / SWITCHES
JUMPER / SWITCH DESCRIPTION
W1 2--1 AUXILARY BUCKY SUPPLY
2--3 NORMAL
EXPOSURE ENABLE RELAY ( K4 ) CAN BE SUPPLIED FROM
INTERNAL +24VDC ( NORMAL ), OR FROM EXTERNAL
VOLTAGE ( AUXILIARY BUCKY SUPPLY ).
W2 2--1 NORMAL
2--3 LOGIC LINE SYNC
FLUORO EXPOSURE SYNCHRONIZATION WITH AC LINE
DIRECTLY ( NORMAL ), OR THRU A LOGIC CIRCUITRY.
W3 A POSITIVE ION CHAMBER SUPPLY
B NEGATIVE PHOTOMULTIPLIER TUBE SUPPLY
W4 A POSITIVE ION CHAMBER SUPPLY
B NEGATIVE PHOTOMULTIPLIER TUBE SUPPLY
W5 A POSITIVE ION CHAMBER SUPPLY
B NEGATIVE PHOTOMULTIPLIER TUBE SUPPLY
W6 A POSITIVE ION CHAMBER SUPPLY
B NEGATIVE PHOTOMULTIPLIER TUBE SUPPLY
THE OUTPUT OF THE PHOTO TUBE/ION CHAMBER HIGH
VOLTAGE POWER SUPPLY CAN BE SELECTED A S :
W7 A POSITIVE ION CHAMBER SUPPLY
B NEGATIVE PHOTOMULTIPLIER TUBE SUPPLY
-- POSITIVE OUTPUT ( J UMPER POSITION -- A )
-- N E G
TI
E OUTPUT ( JUMPER POSITION -- B )
W8 A POSITIVE ION CHAMBER SUPPLY
B NEGATIVE PHOTOMULTIPLIER TUBE SUPPLY
W9 A POSITIVE ION CHAMBER SUPPLY
B NEGATIVE PHOTOMULTIPLIER TUBE SUPPLY
W10 A POSITIVE ION CHAMBER SUPPLY
B NEGATIVE PHOTOMULTIPLIER TUBE SUPPLY