HF Series Generators
Installation
IN-1052R0
26
GENERATOR TOUCH SCREEN (TPC) USING A POSITIONER WITH
TOUCH SCREEN
For Generator using a TPC, perform the following connections:
(Refer to Illustration 3-4)
1. Connect the AC Supply Cable (A7059--xx) from “AC Supply” connector
of the TPC to the respective terminals in TS1 of the Generator Cabinet.
2. Connect the Serial Interconnection Cable ( A3352--xx) from “COM1
(Generator)”oftheTPCtoJ2“Console” of the System Interface Panel
in the Control Box.
3. Connect the Serial Interconnection Cable (A7067--xx) from J1
“Generator” of the System Interface Panel in the Control Box to J5 of the
Generator Cabinet.
4. Connect the Handswitch Cable to the Handswitch connector of the TPC.
5. The TPC includes extra connectors for the Mouse and the Keyboard.
Connect them as indicated in the illustration below. In some cases, these
communication ports are used to interface the TPC with other devices.
For further information, refer to Section 5.2 -- Maps 54301052 and
A6188--03 at the end of this Document, and refer to the respective
Positioner Service Manual.
Note .