O
O
PIN
1
2
3
4
5
6
7
12
13
I/O
O
O
O
O
I
I
O
I
SIGNAL
SUPPLY (BUCKY 1)
BUCKY SPLY 1
0 VAC (BUCKY 1)
BUCKY 1 DR
--BUCKY 1 MOT
BUCKY 1 MOT RTN
GND
IC GND
L2/N
L3
TERMINAL BLOCK 3TS1
COMPACT GENERATOR CABINET
SEDECAL
54301045
DWG:
NAME DATE
DRAWING
REVISED
SHEET / OF
AREVBCDEF
F. GARCIA 04/04/99
1/2
A. DIAZ 28/01/00
REV DESCRIPTION ISSUED BY DATE
TO ROTOR TUBE
F. GARCIA 05/05/00
New interfaceA
F. GARCIA 04/01/01
CN 00/211B
SIGNAL
COM T1
AUX T1
MAIN T1
I/O
I
I
PIN
1
2
3
SIGNAL
GND UNR
+12V UNR
+12VDC
-- P W R O F F
-- P W R O N
GND UNR
GND
+12V UNR
--12VDC
SIGNAL
--EXP
C--HT DAT
HT--C DAT
--PREP
C--HT CLK
HT--C CLK
HV PT CRL
-- K V D W N
LINE SYNC
--LINE CONT
--BUCKY 2 DR CMD
--BUCKY EXP
-- H T I N L
--BUCKY 1 DR CMD
-- K V U P
SIGNAL
-- G E N O K
--SFD SEL
--THERMOSTAT 1
--SF PREP
GND (THERM. COM)
--COLLIMATOR
TABLE ERR(COMP)
--ROOM LIGHT
--READY
EXP OK
ABC OUT/--LEFT
--DIRECT SEL
PT INPUT
EXT SYNC (FL DSI)
SPARE IN2
--SFC (--PT SEL)
--FL EXP
GND (DOOR RTN)
--DOOR
--THERMOSTAT 2
-- A U T O O F F
SPARE IN1
ALOE
-- A L O E
--ACT EXP
I/O
I
I
I
O
O
I
I
I
I
I/O
O
O
I
O
O
I
O
O
I
O
O
I
I
O
O
I/O
O
O
I
I
I/O
I
I
O
O
O
O
O
I
I
I
I
I/O
I
I
O
I
O
O
PIN
1
2
3
4
5
6
7
8
9
PIN
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
PIN
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
INTERFACE CABLE
CONNECTOR J1
CONNECTOR J3
CONNECTOR J2
POWER
COMMUNICATION
INTERFACE
O
SIGNAL
GND
+12V UNR
+12VDC
-- P W R O F F
-- P W R O N
GND
GND
+12V UNR
-- 1 2 V D C
I/O
O
O
O
I
I
O
O
O
O
PIN
1
2
3
4
5
6
7
8
9
CONNECTOR 6J2
ADAPTATIONS
BOARDS
COMMUNICATION CABLE
POWER CABLE
}
TO LINE
TO EARTH GROUND
I
BREAKER
VOLTAGE
SIGNAL
COM
PREP
EXP
I/O
O
I
I
PIN
1
2
3
CONNECTOR J15
HAND--SWITCH (or VET PEDAL SWITCH)
GND
14
16
17
O
O
O
PT SPLY
-- 1 2 V
--ROOM LIGHT
+24V UNR
18 O
15
O+12V
GND
STUD
10
O
ROOM LIGHT SW
23
O
22
I
11
O
ROOM LIGHT SUP
9
O
220 VAC SW
8
O
115 VAC SW
SIGNAL
IC1 INPUT
IC3 INPUT
--FLD1 DR
--FLD3 DR
GND
IC2 INPUT
IC4 INPUT
--FLD2 DR
-- S T R T D R
I/O
I
I
O
O
O
I
I
O
O
PIN
1
2
3
4
5
6
7
8
9
CONNECTOR J5
A.E.C.
O
L1
GND
STUD
GND CABLE
24
26
O
--DOOR
27
39
L3 only for Three Phase
SYSTEM INTERCONNECTION
4N.U.
AEC CABLE
I
O
O
THERMOST. COMM.
--THERMOSTAT 1
O
4
5
O
42
SUPPLY (BUCKY 2)
BUCKY SPLY 2
0 VAC (BUCKY 2)
BUCKY 2 DR
--BUCKY 2 MOT
BUCKY 2 MOT RTN
GND
O
48
47
O
GND
DOOR RTN
O
O
PT INPUT
--FL EXP
--READY
-- A L O E
--SF PREP
36
37
I
51
O
53
52
O
SIGNAL
--4 IN SEL
--9 IN SEL
+12 VDC
V SYNC
-- F T S W C M D
FL START
--CAM FL EXP
N.U.
--ABS
--6 IN SEL
BEEP
GND
-- M E M E N
-- M E M G AT E
N.U.
I/O
O
O
O
I
I
O
O
O
O
O
PIN
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
CONNECTOR J1 / J10
FLUORO CPU BOARD
O
O
O
FLUORO CABLE
SIGNAL
--THERMOSTAT 1
--THERMOSTAT 2
THERM. COMM
I/O
I
I
I
PIN
1
2
3
TERMINAL BLOCK 4TS3
COMPACT / COMPACT-ESM GENERATORS
FAN 1
O
O
6
O
GND
0VAC
7
8
COM T2
AUX T2
MAIN T2
I
I
9
10
11
O
THERMOST. COMM.
--THERMOSTAT 2
O
12
13
O
FAN 2
O
O
14
O
GND
0VAC
15
16
COLLIMATOR LAMP
SIGNAL
+24 VDC LOCKS
0VACLAMP
24 VAC LAMP
I/O
O
O
PIN
3
4
5
O
TERMINAL BLOCK TB7
+24 VDC LOCKS
0 VDC LOCKS
O
6
7
O
0 VDC LOCKS
8
O
LOCKS BOARD
}
LOCKS
TABLE ERROR
--SFC (--PT SEL)
I
I
5
7
-- G E N O K
--SFD SEL
--DIRECT SEL
EXP OK
O
O
O
O
10
11
12
15
--ACT EXPO
16
-- A U T O O F F
O
17
SIGNAL
--EXP
C--HT DAT
HT--C DAT
--PREP
C--HT CLK
HT--C CLK
HV PT CRL
-- K V D W N
LINE SYNC
--LINE CONT
--BUCKY 2 DR CMD
--BUCKY EXP
-- H T I N L
--BUCKY 1 DR CMD
-- K V U P
I/O
O
O
I
O
O
I
O
O
I
O
O
I
I
O
O
PIN
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
CONNECTOR 6J3
GNDO16
F. GARCIA 02/02/01
New interfaceC
TERMINAL BLOCK
The ROTOR TUBE connections
are made to TS2 on the DRAC
NOTE.-- For Generator with DRAC :
TS2-- 6 is GND when
TS2 only has 6 terminals.
NOTE :
NOTE -- For Serial Cabinet :
-- the Interface cable connections are factory
made to 3TS1, TS2 and 4TS3 terminal blocks.
-- 6J2 is not supplied, Power cable is directly
connected in factory to J1 of the ATP Console.
-- 6J3 is not supplied, Power cable is directly
connected in factory to J3 of the ATP Console.
4TS2 / 10TS2 / 11TS2
SIGNAL
TIME1
TIME2
TIME3
TIME4
--PS (DSI) SEL
--CINE (DSA) SEL
--HCF SEL
SPARE IN
--TOMO PREP
--TOMO EXP
TOMO ON
EXT REF
GND
EXP STOP
-- F S
I/O
I
I
I
I
I
I
O
O
O
I
I/O
I
PIN
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
CONNECTOR J13
TOMO
I
O
I
TOMO CABLE
(Table & Tube Stand)
F. GARCIA 08/01/02
New interfaceD
FOR GENERATORS
LINE POWERED
HAND--SWITCH
(or VET PEDAL SWITCH)
SIGNAL
C--FL DAT
--CAM FL EXP
FL--C DAT
FL-- C CLK
+12V ISO
GND
C--FL CLK
GND
CAM SYNC
I/O
O
I
I
I
I
I/O
I
I/O
O
PIN
1
2
3
4
5
6
7
8
9
CONNECTOR J4
SIGNAL
C--FL DAT
GND
--CAM FL EXP
C--FL CLK
FL--C DAT
GND
FL-- C CLK
CAM SYNC
+12V ISO
N.U.
I/O
I
I/O
O
I
O
O
I
PIN
1
2
3
4
5
6
7
8
9
10
CONNECTOR J4
I/O
O
ALOE
O
8
ABC OUT / --LEFT
I
18
EXT SYNC(FL DSI)
I
19
-- C O L L I M AT O R
I
20
ATP CONSOLE BOARD
FLUORO
Signal for Thermostats go to 4TS3
andthentoTS2orgodirectlytoTS2
depending on Generator model
NOTE :
2
3
1
FOR GENERATORS
BATTERIES POWERED
1TB1--LINE
}
TO SOCKET
(Line Voltage)
N
GND
L
Foot Switch Cmd.I
A
GND
B
Prep/Rdy Acq Rad
O
C
Boost Fluoro
I
D
Digital Exposure
I
E
F. GARCIA 26/04/04
E
CN 04/032
54 O 220 VAC SPLY
F. GARCIA 01/09/06
F
Adaptation Boards