Spectroquant
®
Pharo 300 Operation
85
ba75703d07 04/2014
3 In the Target value column, enter
the nominal values of the individ-
ual standard solutions.
The nominal value for E0 (reagent
blank value) is preset and cannot
be changed. The respective
absorbance has to be measured.
4 Select an absorbance value and
confirm with <START ENTER>.
The measurement window pops
up.
5 Insert the cell with the relevant
standard or the reagent blank
value (for E0).
The first single measurement for
the calibration is carried out.
The following data is displayed as
the result:
The measured absorbance
from the (last) single measure-
ment.
The median from all single mea-
surements carried out up to
now.
6 If necessary, carry out further sin-
gle measurements for the forma-
tion of the median with [Next
meas.]
or
discard the last single measure-
ment with [Discard].
7 To accept the median value, press
[Apply].
The list of the standards required
for this method pops up. The
absorbance measured for the
standard or reagent blank value
respectively (E0) is entered.
Calibrate the method
16.01.12 9:52
Target value (Ca) Absorbance
E0 0.00 mg/l
10.60 mg/l
21.50 mg/l
32.40 mg/l
43.20 mg/l
54.00 mg/l
Back Next
Calibrate the method
16.01.12 9:52
To start measurement, insert cell or press
<START/ENTER>
304: Ca
Ca
10 mm
0..20 - 4.00 mg/l
Calibrate the method
16.01.12 9:52
Last measured absorbance
0.177
Median
0.177 (1 Measurement(s))
304: Ca
Ca
10 mm
0..20 - 4.00 mg/l
Next meas.
Discard Apply