Analyzing waveforms Set up MIPI search parameters
Behavior
Clicking an entry in the Type column activates the individual search controls, similar to the one shown
above. These controls vary depending on the search Type. Setting the search conditions is similar to
setting the parameters for an Event Trigger
(see page 386).
NOTE. No info
rmation will be displayed in this window until you specify at least one search Type in
the Search Select Control Window (see page 300).
To use
Click the Source box and select the bus number or name from the drop-down list.
Select the type of bus cycle o r activity to search for from the Search For drop-down list.
For some S
earch For selections, you must set additional fields to define other parameters such as for
aLongPacket.
In this example, you need to set a Data Type Value, Virtual Channel, Direction, Word Count, Number of
Bytes, Payload Format, Payload Value, and Byte Offset (click the Edit
(see page 939) button). Optionally,
you can also set the component threshold levels through the Logic Thresholds Setup button.
Click the Search For selections Stop (see page 338), Start of Transmission (see page 338), End of
Transmission (see page 338), Bus Turn Around (see page 339), Escape Mode (see page 339), Warning
(see page 339), Error (see page 340), Short Packet (see page 341),orLong Packet (see page 342) to
view their parameters.
What do you want to do next?
Learn about bus setups. (see page 95)
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 337