7 Maintaining the product
0970 4020 en 01 145
Saveris T3/T3D radio probe
2-channel radio probe with 2 external TC probe connections
(TC characteristics can be selected)
Characteristic Values
Probe type TC
Measuring range
TC type J
TC type K
TC type S
TC type T
-100 to +750 °C
-195 to +1350 °C
0 to +1760 °C
-200 to +400 °C
Accuracy ±0.5 °C or 0.5 % of measured value
Resolution 0.1 °C/TC type S 1 °C
Connection
2 TCs via TC socket,
max. difference in potential 2 V
Protection class
IP54 (with probe inserted or port sealed
with rubber plug)
Operating
temperature
-20 to +50 °C
The probe inputs are not isolated from one another.
Take this into account when using probes with non-
isolated thermocouple.

Table of Contents

Other manuals for TESTO Saveris

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO Saveris and is the answer not in the manual?

TESTO Saveris Specifications

General IconGeneral
Humidity Range0 to 100 %RH
Measurement ParametersTemperature, Humidity
Data TransmissionWireless
Alarm FunctionYes
Resolution0.1 °C, 0.1 %RH
Radio Frequency868 MHz, 915 MHz
Data StorageInternal memory, Cloud
Power SupplyBattery