EasyManua.ls Logo

Topcon SR-LEDW - Output Format for Reading out the Internal Stored Data

Topcon SR-LEDW
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 98 -
4.2.2
Output Format for Reading out the Internal Stored Data
Output format of the data which was measured in the local mode and stored in the SR-LEDW/SR-
UL2/SR-UL1R/SR-3AR the measured data is returned with text from the instrument.
Line No. Output data Remarks
1. n Data number
2. 2 Measuring field (degree)
3. 100 Integral time (milli-second)
4. 9.335E-01 Radiance (W•m
-2
•sr
-1
)
5. 1.490E+02 Luminance (cd/m
2
)
6. 1.631E+02 Tristimulus values X
7. 1.490E+02 Tristimulus values Y
8. 5.374E+01 Tristimulus values Z
9. 0.4458 Chromaticity x
10. 0.4073 Chromaticity y
11. 0.2549 Chromaticity u’
12. 0.5240 Chromaticity v’
13. 2882 Correlated color temperature (K)
14. 0.0002 Deviation
15. 380 2.141231E-04 Wavelength (nm), Spectral radiance (W•m
-2
•nm
-1
•sr
-1
)
16. 381 2.420037E-04
・・・・・・・・・・・・・・・・・・・
414. 779 4.325765E-03
415. 780 4.294558E-03
416. END Data end command

Table of Contents

Related product manuals