- 46 -
3.6.1 High Speed Calibration
Performe High Speed Calibration when select HIGH SPEED or HIGH SPEED 2, HIGH SPEED 3.
Use the following procedure to set the measuring speed;.
1 Switch to the function mode to display the
[MEASURE] -[High Speed Cal]
'3.1.2 Entering/Returning from the Function Mode'
*High Spee Cal.*
CHANGE: Calibration
2
Press the [CHANGE] switch.
*High Speed Cal.*
SHIFT : OK
CHANGE : NO
3 The calibration starts after closing finder shutter and pressing the [SHIFT] switch.
It takes about three minute to complete calibration
'1.3.5 High Speed Calibration'
Remember!
Close the finder shutter before calibration.