- 58 -
3.15 Measurement Averaging
Set whether to average measurements.
Normal Meas : Do not use averaged measurements.
Average Meas : Use averaged measurements.
Measurement averaging allows the average value over multiple measurements to be set as the
measurement value to increase precision for measurements of light sources with low luminance.
Note that the higher the number of measurements is used, the longer time the measurement process
will take.
Measurement time = Integral time * (Average count + 1) + Integral time + Shutter operation time +
Processing time
Setting whether average measurement is used or not
1 Switch to the function mode to display the [MEASURE]-
[AverageMeas ON/OFF].
☞
'3.1.2 Entering/Returning from the Function Mode'
*AverageMeas O N/OFF*
*Normal Meas
2 To change the setting, press the [CHANGE] switch.
Press the [ROTATION] switch to select the desired value.
The value cycles as follows:
Normal Meas → Average Meas
*AverageMeas ON/OFF*
*Average Meas
When remote mode measurement and measurement mode are MANU mode, averaged
measurement is disabled.