- 84 -
4.1.20 FLD1, FLD2, FLD3 and FLD4 Command
Changes the measuring field setting in the SR-LEDW/SR-UL2/SR-UL1R/SR-3AR.
FLD1: Change the measuring field to the 2.0 position.
FLD2: Change the measuring field to the 1.0 position.
FLD3: Change the measuring field to the 0.2 position.
FLD4: Change the measuring field to the 0.1 position.
PC Command communication direction Measuring instrument
FLD1/FLD2/FLD3/FLD4+(CrLf)
OK+(CrLf)
4.1.21 FLDR Command
Reads the current measuring field setting from the SR-LEDW/SR-UL2/SR-UL1R/SR-3AR.
PC Command communication direction Measuring instrument
FLDR+(CrLf)
OK+(CrLf)
setting+(CrLf)
END+(CrLf)
4.1.22 NL Command
Sets the integral time for the integral time delay function in the SR-LEDW/SR-UL2/SR-UL1R/SR-
3AR.
NL_####The sign (#) is the integral time from 50 to 3 000 ms.
PC Command communication direction Measuring instrument
NL+(CrLf)
OK+(CrLf)