EasyManua.ls Logo

Topcon SR-LEDW - Dr[N] Command; KO1;KO2;KN1;KN2 Command

Topcon SR-LEDW
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 80 -
4.1.11 DR[n] Command
The local measurement is completed and the measurement data stored by the instrument is read
out by the computer.
“n” is specified with 1 ~ 50.
‘1.3.2 Connecting the Personal Computer’
‘4.2.2 Output Format for Reading out the Internal Stored Data’
PC Command communication direction Measuring instrument
DRn + (CrLf)
OK+(Cr•Lf)
Data No. + (CrLf)
380+(Cr•Lf)
• • • • • • •
780... + (CrLf)
END +(Cr•Lf)
4.1.12 KO1/KO2/KN1/KN2 Command
Change the use of the correction factors in the instrument.
KO1 : Changes the setting to use the correction factor for spectrum in the instrument.
KO2 : Changes the setting to use the correction factor for tristimulus values in the instrument.
KN1 : Changes the setting not to use the correction factors for spectrum in the instrument.
KN2 : Changes the setting not to use the correction factors for tristimulus values in the
instrument.
PC Command communication direction Measuring instrument
KO1/KO2/KN1/KN2+(CrLf)
OK+(CrLf)

Table of Contents

Related product manuals