- 42 -
3.4 Integral Time Delay Function
Set whether to use the integral time delay function or not. Explain how to set the integral time delay
function to usable state as follows.
The followings cases will result in measurement errors.
1. When measuring a light source with high duty ratio as well as high intensity
2. When periodically-changing light sources include very low intensity such as black
3. When cycles of a light source which are measured within the given integral time are not enough
Increasing the integral time is effective in decreasing the error. The integral time delay function
can be used to increase the integration time, to allow stable measurements.
'2.5.4 Measuring a Frequency Light Source'
Use the following procedure to set the integral time delay:
1 Switch to the function mode to display [MEASURE]-
[Integ Delay ON/OFF].
'3.1.2 Entering/Returning from the Function Mode'
2 Press the [CHANGE] switch.
3 To change the setting, press the [ROTATION] switch. ON/OFF is changed to each other.
*Integ Delay ON/OFF
*ON
4 Press the [ENTER] switch to accept the displayed value.
DIF.
ABS.
UP
LAMP
DOWN
MEAS.
HOLD
FIELD
FUNCTION
ENTER
CHANGE
ROTATION
SHIFT
DIF.
ABS.
UP
LAMP
DOWN
MEAS.
HOLD
FIELD
FUNCTION
ENTER
CHANGE
ROTATION
SHIFT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon SR-LEDW and is the answer not in the manual?

Topcon SR-LEDW Specifications

General IconGeneral
BrandTopcon
ModelSR-LEDW
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals