4.1.8 KR[n] Command
The correction factors stored by the instrument are read out. One of 0~400 is inputted to “n” and it
corresponds to every 1nm interval in the range of 380 nm~780 nm.
KR0 : 380 nm
KR1 : 381 nm
KR2 : 382 nm
• • • • • •
KR399 : 779 nm
KR400 : 780 nm
PC Command communication direction Measuring instrument
KRn + (CrLf)
OK+(Cr•Lf)
Correction factor + (CrLf)
• • • • • • •
KRn+(Cr•Lf)
OK + (CrLf)
Correction factor +(Cr•Lf)
4.1.9 KX, KY and KZ Command
With these commands, the correction factors for tristimulus values XYZ are stored in the instrument.
KX_##### : "_" is space. "#####" is correction factor.
KY_##### : "_" is space. "#####" is correction factor.
KZ_##### : "_" is space. "#####" is correction factor.
PC Command communication direction Measuring instrument
"KX", KY or "KZ" + (CrLf)
OK+(Cr•Lf)