- 87 -
4.1.28 FX Command
Starts the measurement to calculate the optimum inner-filter and integral time setting in FIX INTEG
mode. When the measurement complete, the optimum inner-filter type and integral time is retuned
and the values are also set on the instrument. (SR-LEDW only)
PC Command communication direction Measuring instrument
FX + (Cr・Lf)
OK + (Cr・Lf)
Integral time+ (Cr・Lf)
Inner-filter+ (Cr・Lf)
END” + (Cr・Lf)
This measurement should be conducted at the same condition as the actual
measurement in measurement angle, object and measurement distance.
The integral time and inner-filter setting in FIX INTEG mode are erased
when the power off.
4.1.29 FO Command
Sets to FIX INTEG mode(SR-LEDW only)
This setting is stored even after the power off
PC Command communication direction Measuring instrument