- 125 -
Observation visual field 10
samp
2
1
10
10
LxKX
2
1
samp
10
10
LyKY
2
1
samp
10
10
LzKZ
¯x10 (), ¯y10 (), ¯z10 (): Color matching functions in CIE 1964 auxiliary standard colorimetric
system
K: Parameter 683 lm/W
1: Measuring start wavelength 380 nm
2: Measuring end wavelength 780 nm
: =1
* “Y” of the observation visual field 2 is used for the luminance Lv of the observation visual field
10.
Chromaticity coordinates
Observation visual field 2
“xy” chromaticity coordinates in XYZ colorimetric system
ZYX
X
x
ZYX
Y
y
“u’v’” chromaticity coordinates in UCS colorimetric system
3Z15YX
4X
u'
3Z15YX
9Y
v'
Observation visual field 10
“x
10
y
10
” chromaticity coordinates in XYZ colorimetric system
101010
10
10
ZYX
X
x
101010
10
10
ZYX
Y
y
“u’v’” chromaticity coordinates in UCS colorimetric system
101010
10
10
3Z15YX
4X
u'
101010
10
10
3Z15YX
9Y
'v
Correlated color temperature and deviation
To calculate the correlated color temperature and deviation, use the spectral distribution of the JIS Z
8725 light source and the measurement method of the correlated color temperature.
Correlated color temperature display range: 1563 KTc100 000 K
Deviation display range: -0.02duv0.02
Deviation means the distance from the black body radiation locus on CIE 1960 UCS chromaticity
diagram.