157
MAINTENANCE AND CHECKING
4 Wipe the glass surface lightly with the applicator, from the center outward.
5 Use a new applicator and wipe the glass surface repeatedly several times.
6 Cleaning is completed when oil spots are removed.
If stains cannot be removed easily, call your dealer.
CLEANING THE NOZZLE AND CHAMBER GLASS INSIDE THE MEASURING NOZZLE
• If there is any foreign matter on or around the measuring nozzle, it may enter and damage
the patient's eye during the measurement. If any, clean the nozzle.
• When the window glass inside the nozzle becomes stained, it makes the fixation target
unclear, causing errors in auto alignment and measurement values. If the fixation target is
unclear or measurement values with parentheses are frequent, clean the window glass
inside the nozzle.
• Clean the glass when "CLEAN THE CHAMBER GLASS" is displayed on the monitor
screen.
1 Prepare the ethanol.
2 Moisten the applicator with ethanol.
Repeat the cleaning operation several times, using a new applicator each time, till
oil spots are removed from the window glass.
When the measuring window glass becomes stained "CLEAN THE MEASURING
WINDOW GLASS" is displayed on the monitor screen.
CAUTION
Clean the measuring window glass, measuring nozzle and the
window glass inside the measuring nozzle using ethanol.
Other chemicals may cause damage to the patient's eye during
measurement.
MEMO
• Do not apply unreasonable force to the measuring nozzle while clean-
ing.
• To avoid problems, do not leave the cotton fibers inside.
• Be sure to use only the attached applicator.
Wiping the glass surface