MESSAGE LIST
"OVER-SPH" Spherical power exceeds +22D or -25D (at VD=12mm).
"OVER-CYL" Cylindrical power exceeds ±8D (at VD=12mm).
"OVER-R" Corneal curvature exceeds 5.00-10.00mm.
"NO TARGET" No patient's eye or the eye image is too dark.
"AGAIN" A difference of more than 5D from the value of previous measure-
ment
"NO CENTER" Not possible to find the center of patient's eye.
"ALIGN ERROR" In measuring, aligning fails.
"PAPER END" Printer paper is used up.
"CLOSE PRINTER COVER"
Printer cover is open when PRINT button is pressed.
"PRINT" Printing is under way.
"ERROR" Patient's eye blinked or moved during measurement.
If this appears when measurement is done properly with the
model eye, a problem is suspected: Call our service engineer.
"CUTTER INIT ERR" The position of the printer auto cutter is incorrect.
"PUSH PRINT SW" Indicates printing steps under no paper/head up condition.
"NO ID" ID input is not done under ID input setting.
"RS232C DATAOUT" RS232C data transmission is under way.
"RS232C SUCCESS" RS232C data transmission is completed.
"RS232C FAIL" RS232C data transmission has failed.
"CABLE UNCONNECTED" The RS232C cable is not connected, or the power supply of the
other party is OFF.
"WAITING" Ready for RS232C transmission (CTS), and waiting for the signal.
"ERROR" Error in RS232C data input.
"READY" Ready for RS232C data input.
"USB DATAOUT" USB data transmission is under way.
"USB SUCCESS" USB data transmission is completed.
"USB FAIL" USB data transmission has failed.
"NOZZLE LIMIT" The main body exceeds the limit to the patient side.