6.  Disconnect the power and replace the communication modules. 
-  Replace one module at a time and try to restart the system to 
find out the damaged one. 
-  Restart the system. 
7.  The QML logger does not receive commands entered in the AWS 
Client software. 
-  Check the maintenance cable connection and that you have the 
correct maintenance cable in use. 
-  The terminal connection is not open. 
- Type open and press ENTER. Note that the command must be 
typed exactly correctly before it can be executed and that the 
command is not echoed on the screen. 
-  The port settings on PC and QML logger are not matching 
-  Use settings 9600bps, 8 data, 1 stop, no parity on PC. 
8.  Has someone made any software or hardware changes to the station 
prior to malfunction? If so, could those changes have an effect on 
the operation? 
a.  When you have loaded a new setup: 
-  Make sure you have the original setup and the new setup 
saved as an .adc file. 
-  Reload the original setup and verify that the system starts 
and runs with the original setup. 
b.  When you have installed new hardware: 
-  Disconnect the new hardware. 
-  Reload the original setup and verify that the system starts 
and runs. 
-  Connect the new hardware. 
-  Reload the new setup and verify that the system starts and 
runs. 
9.  Sensor(s) are not working properly. 
a.  Are there one or multiple malfunctioning sensors? Multiple 
simultaneous sensor malfunctions in the same QML logger 
often indicate a damaged logger. 
b.  Check the output of the warnings and errors commands for 
indication of the probable cause. 
-  Connect the maintenance cable and start the AWS Client 
software. Type commands warnings and errors. For 
details, see section Warnings and Errors on page 219. 
-  Pay also attention to the message related to the excitation 
voltages. If the damaged sensor is powered by the logger, 
it may cause an overload.