User's Guide ______________________________________________________________________  
212 _________________________________________________________________ M211296EN-A 
Resetting QML Logger 
To reset the QML logger, enter the reset command (recommended), or 
press the reset button indicated by number 1 in Figure 121 below. 
 
  1004-121 
Figure 121  Reset Button on QML Logger 
A short reset (pressing the reset button quickly) performs the same reset 
as giving the command and starts the program again. A long reset 
(pressing the reset button and keeping it down for a few seconds) restarts 
the program with a so-called blank setup. A blank setup does not run a 
setup file. 
A blank setup may be useful if the setup is somehow defective and does 
not allow the opening of a terminal connection. When the blank setup is 
run, the QML logger communication parameters are restored to their 
defaults: 9600bps, 8 data, No parity, 1 stop bit.