Model Description
‑2
4VDC With electrical installation 24 VDC, PR‑9298
1) Steam connection: Tri-Clamp DIN10 DIN32676.
10.4.3 PR‑23‑AP specifica
tions
Table 26 General specifica
tions
Feature Specification
Refractive index range Full range n
D
1.
3200 … 1.5300 (corresponds to hot
water – 100 Brix)
Accuracy Refractive index n
D
±0
.0002 (corresponds typically to
±0.1 % by weight)
Repeatability and stability correspond to accuracy
Speed of response 1 s undamped, damping time selectable up to 5 min
Calibration With Cargille c
ertified refractive index liquids over full
range of n
D
1.3200 … 1.5300
CORE‑Optics No mechanical adjustments
Digital measurement 3648 pixel CCD element
Light source LED 589 nm wavelength, sodium light
Temperature sensor Built‑in P
t‑1000
Temperature compensation Automatic, digital compensation
Instrument v
erification With certified refractive index liquids and Vaisala
documented procedure
Ambient temperature −20 °C … 45 °C (−4 °F … 113 °F)
Table 27 Sensor PR‑23‑AP specifica
tions
Feature Specification
Sensor PR‑23‑AP Probe sensor model for large pipe lines and vessels
Process connection Sanitary clamp 2.5 in; sanitary clamp 4 in or MT4 DN
25
/1T APV Tank bottom flange
Sanitary design 3‑A S
anitary Standard 46‑04 approved and EHEDG
(European Hygienic Engineering & Design Group)
certified (connections ‑T, ‑N, ‑R, ‑S, ‑H, ‑C)
Process pressure Sanitary clamp max. 15 bar (200 psi) at 20 °C
(7
0 °F) / 9 bar (125 psi) at 120 °C (250 °F)
Process temperature −20 °C … +150 °C (−4 °F … +302 °F)
Process wetted parts, standard AISI 316L stainless steel, spinel or sapphire prism,
prism gask
ets MTF (Modified Teflonâ)
PR-23 Series User Guide
IM-EN-PR23-E
118