EasyManua.ls Logo

Vaisala PR-23 Series - Figure 39 Concentration Calibration Layers

Vaisala PR-23 Series
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
mA OUTPUT
FIELD CALIBRATION
CHEMICAL CURVE
n
D
CCD
TEMP
Pt-1000
Indicating
transmitter
DTR
Sensor
CONC
CALC
1.
2.
3.
4.
5.
6.
DAMPING
Figure 39 Concentration calibration layers
1 The information from the CCD element and the Pt-1000 temperature element. The
position of the shado
w edge, see Figure 143 (page 259), is described by a number called
CCD and scaled 0 … 100 %.
2 Sensor calibration: The actual refractive index n
D
is calculated from the CCD value. The
process temperature is calculated from the Pt-1000 resistance. The sensor output is n
D
and temperature TEMP in degrees Celsius. The calibrations of all sensors are identical,
which makes sensors interchangeable. The calibration of each sensor can be verified
using standard refractive index liquids.
3 Chemical curve: The indicating transmitter DTR receives n
D
and TEMP and calculates the
concentration value according to chemical curves. The result is a temperature-
compensated calculated concentration value CALC.
Chapter 7 – C
onfiguration and calibration
73

Table of Contents

Related product manuals