EasyManuals Logo

Vaisala PR-23 Series User Manual

Vaisala PR-23 Series
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
3.3 Pipe mounting checklist
Most in-line refractometer models are mounted in a pipe. The recommended minimum flow
velocity is 1.5 m/s (5 ft/s). The diameter and form of the pipe and the process temperature
aect
the measurement and need to be taken into account.
1. If the process pipe diameter varies, select the position with the smallest diameter (and
accordingly highest velocity). Then the prism keeps better clean.
2. If the refractometer is used in a feed-back control loop, make the time lag short. When
a dilution valve is controlled, for example, mount the refractometer close to the dilution
point. However, make sure complete mixing has occurred at mounting location.
3. If the temperature varies along the process pipe, select the position with the highest
process temperature. Then the risk of prism coating is minimized, because higher
temperature means higher solubility and also lower viscosity.
4. Often the position with the highest process pressure (= after pump + before valve) has
favorable flow
conditions without sedimentation or air trapping risks.
5. The sensor is accessible for service.
3.4 Checklist for mounting in tank, vessel
or large pipe
A probe sensor PR-23-AP or PR-23-GP can be inserted with a flange or clamp into tanks and
vessels which either do not have a scraper or where the mixer does not touch the vessel
wall. A probe sensor can also be flush mounted in a cooker where the scraper touches the
wall.
1. The inserted probe sensor is mounted close to a stirrer to ensure representative sample
of the process liquid and to keep the prism clean.
2. The sensor is accessible for service.
PR-23 Series User Guide
IM-EN-PR23-E
20

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala PR-23 Series and is the answer not in the manual?

Vaisala PR-23 Series Specifications

General IconGeneral
BrandVaisala
ModelPR-23 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals